Mini-Oral Presentation

Luminex mixed bead and single antigen bead assays have high positive percentage agreement for HLA-A, B and DR, with less agreement identified at HLA-DQ

Presenter
Petrina Guthrie (AU)

Australian Red Cross Lifeblood ยท Alexandria

Profile

Authors

Petrina Guthrie1, Daniel Russell1, Su Hyun Lee1, Robert Carroll1, Gerard Chu1

Affiliations

1 Australian Red Cross Lifeblood

Abstract

Introduction: The Luminex Labscreen mixed (LM) assay is a screening test to detect anti human leucocyte antigen (HLA) antibodies. However, the LM may be less sensitive than Luminex Single Antigen Bead (SAB) technology currently used to screen for anti-HLA antibodies in Australia. Testing algorithms assess accuracy and cost-effectiveness to facilitate safe transplantation in an economically feasible manner. One potential algorithm in Australia is testing unsensitised kidney recipients with prior negative SAB results with periodic LM assays. We compared the LM with SAB to determine whether the LM assay was a sensitive screening alternative to identify anti-HLA antibodies at clinically relevant levels.
Methods: Serum samples from 45 patients with anti-HLA-A, B, DR and DQ antibodies detected by One Lambda SAB were selected for analysis. Specimens were chosen based on mean fluorescence intensity (MFI) at 2000-4000 and 4000-6000 because these define unacceptable antigens for living and deceased donor kidney programs in Australia. Samples were chosen based on single-locus reactivity to allow locus-by-locus comparison, and profiles were chosen to target known serological targets or antibody-verified Eplets. Unsensitised males and sera with known non-specific reactivity patterns were excluded. Samples were then tested by the LM assay. A normalised background ratio of 2.1-2.4 (borderline) or ≥2.4 (positive) on LM was considered a positive LM screen. Discrepant results were tested using absorption with crossmatch cells and elution (AXE) to confirm real reactivity. True positives were defined as positive results on both SAB and LM. When discrepancies existed, a true positive was defined by a positive AXE. Positive percentage agreement (PPA = true positives/[true positives + false negatives]) for the LM was calculated at each locus.
Results: At the A-locus there was 10/11 concordance between SAB and LM. The discordant sample was found by AXE against an HLA-A*29 cell to be negative. At both B and DR loci, there was 10/10 concordance between SAB and LM. At the DQ locus, there was 7/14 concordance between SAB and LM. Discordant samples were found by AXE to be negative in 4 samples (DQ2, DQ2, 87Y and 87F eplets) and positive in 3 samples. The 3 true positives confirmed by AXE (not detected by LM) were antibodies against DQ7 and DQ6 (both at 2000-4000 MFI), and the 52PQ eplet (4000-6000 MFI). There was 100% PPA for the HLA-A, B and DR loci. There was a 70% PPA for HLA-DQ.
Conclusion: The LM is a sensitive screening tool to detect antibodies against anti-HLA- A, B and DR identified by SAB at MFIs relevant to the Australian kidney transplant program. However, it is only moderately sensitive in identifying HLA-DQ antibodies, especially at lower SAB MFI (2000-4000). We hypothesise that this is due to differing HLA-DQ density on the LM and SAB beads. Further evaluation of LM in an unselected populations is underway to clarify other diagnostic performance characteristics.